8.5 Exceptions and Error Conditions
The following exceptions and error conditions can occur with the
LOG SELECT command.
Illegal Value in LOG SELECT Data
If one of the bytes sent with the LOG SELECT data contains an illegal
value or if a reserved bit is set, the CTS returns Check Condition
status. To determine which byte contains the error, issue a REQUEST
SENSE command. The sense key will be set to Illegal Request (5h) and
the Fault Symptom Code field (byte 28) will indicate which of the
consecutively received bytes is in error. For example, if the Page Code
field (byte 00, bits 5 through 0) in the Parameter List Header is
invalid, the value for the Fault Symptom Code byte would be 01h.
Note: The CTS counts the first byte received as byte 1, not as byte 0.
Other Errors
The CTS may return the Fault Symptom Codes (FSCs) listed in
Table 8-4 when thresholds are implemented.
Important
If you want the CTS to return these Fault Symptom Codes to
notify you about changes to the counters, first issue a MODE
SELECT (15h) command and send the Control Mode page
(Page Code=0Ah) with the RLEC bit set to 1. Refer to
Chapter 10 for more information.
In addition, be sure to set the DU bit to 0 and the ETC bit to 1
for the log parameter.